FDA Adverse Event
Malfunction
Summary report: N
R SERIES DEFIBRILLATOR
MDR report key: 4841909
·
Received June 8, 2015
Report
- Report Number
- 1220908-2015-01470
- Event Type
- Malfunction
- Date Received
- June 8, 2015
- Report Date
- May 21, 2015
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K060559
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORP; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY SOLDER JOINT ON A TRANSFORMER ON THE PACE/DEFIB ENGINE ASSEMBLY. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE DISPLAYED "DEFIB FAULT 76" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 369203 | R SERIES DEFIBRILLATOR | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | R SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |