FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 4841898 · Received June 8, 2015

Report

Report Number
1220908-2015-01416
Event Type
Malfunction
Date Received
June 8, 2015
Report Date
May 19, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K972241
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORP; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY TRANSFORMER ON THE SYSTEM BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368091 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA