FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 4841803 · Received June 8, 2015

Report

Report Number
1220908-2015-01447
Event Type
Malfunction
Date Received
June 8, 2015
Report Date
May 19, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K972241
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND THE BRIDGE BOARD WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367835 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION M SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA