FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4841401 · Received June 12, 2015

Report

Report Number
2531779-2015-19622
Event Type
Malfunction
Date Received
June 12, 2015
Report Date
May 28, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 08/04/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2015 WITH THE FOLLOWING FINDINGS: ON INVESTIGATION, THE DISPLAY SCREEN WAS CONFIRMED TO BE DIM/FADED/DISCOLORED. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY COMPARTMENT WAS NOTED TO BE CRACKED FROM THE BUMPER TO THE CASE SEAL.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385212 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1