ANIMAS VIBE
Report
- Report Number
- 2531779-2015-19622
- Event Type
- Malfunction
- Date Received
- June 12, 2015
- Report Date
- May 28, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1: DATE OF SUBMISSION 08/04/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2015 WITH THE FOLLOWING FINDINGS: ON INVESTIGATION, THE DISPLAY SCREEN WAS CONFIRMED TO BE DIM/FADED/DISCOLORED. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY COMPARTMENT WAS NOTED TO BE CRACKED FROM THE BUMPER TO THE CASE SEAL.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 385212 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |