FDA Adverse Event Malfunction Summary report: N

REVOLUTION CEMENT GUN

MDR report key: 4840584 · Received June 12, 2015

Report

Report Number
0001811755-2015-02109
Event Type
Malfunction
Date Received
June 12, 2015
Date of Event
May 15, 2015
Report Date
May 15, 2015
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
JDZ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS NOT DUPLICATED; HOWEVER IT WAS CONFIRMED THROUGH COMPONENT INSPECTION. BASED ON A REVIEW OF THE RISK DOCUMENTATION, A POTENTIAL CAUSE FOR THE RAM BUSHINGS PRODUCING METALS SHAVINGS IS WEAR AT THE BUSHING/ROD INTERFACE. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RECEIVED FOR EVALUATION AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE REVOLUTION CEMENT GUN WAS PRODUCING METAL SHAVINGS DURING CLEANING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT, AND THERE WAS NO ASSOCIATED PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE REVOLUTION CEMENT GUN WAS PRODUCING METAL SHAVINGS DURING CLEANING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT, AND THERE WAS NO ASSOCIATED PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385151 REVOLUTION CEMENT GUN CEMENT, MIXER, FOR CLINICAL USE JDZ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1