FDA Adverse Event Injury Summary report: N

HYDROVIEW INTRAOCULAR LENS

MDR report key: 4840261 · Received June 10, 2015

Report

Report Number
1313525-2015-01572
Event Type
Injury
Date Received
June 10, 2015
Report Date
January 14, 2015
Manufacturer
BAUSCH & LOMB
Product Code
HQL
PMA / PMN Number
P990014
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

HYDROVIEW IOL WAS DISCONTINUED AND IS NO LONGER MANUFACTURED BY B&L. DESPITE ATTEMPTS TO OBTAIN THE LENS LOT NUMBER, IT WAS NOT PROVIDED. THEREFORE, A REVIEW OF THE LOT HISTORY RECORD COULD NOT BE PERFORMED. VISUAL INSPECTION OF THE RETURNED LENS FOUND THE OPTIC NEARLY CUT IN HALF AND BOTH HAPTICS ATTACHED AND NOT DAMAGED. THE OPTIC HAS A CLOUDY WHITE APPEARANCE AND SOME SURFACE AREAS WITH AN ORANGE PEEL APPEARANCE. AN ALIZARIN RED S CHEMICAL STAIN TEST WAS PERFORMED TO DETERMINE IF CALCIUM IS PRESENT ON THE LENS. LARGE AREAS OF THE OPTIC DID TURN RED CONFIRMING CALCIUM. THE CAUSE OF THE CALCIFICATION PHENOMENON IS MULTI-FACTORIAL. THREE MAJOR TYPES OF CALCIFICATION HAVE BEEN PREVIOUSLY DESCRIBED: THE PRIMARY FORM REFERS TO CALCIFICATION DUE TO TISSUES INHERENT TO THE IOL. THE SECONDARY FORM REFERS TO DEPOSITION OF CALCIUM ONTO THE SURFACE OF THE IOL MOST LIKELY DUE TO ENVIRONMENTAL CIRCUMSTANCES (E.G. CHANGES IN THE AQUEOUS SURROUNDING THE IMPLANTED IOL ASSOCIATED WITH PRE-EXISTING OR CONCURRENT DISEASES). BY DEFINITION, IT IS NOT RELATED TO ANY PROBLEM WITH THE IOL ITSELF. THE THIRD FORM IS A FALSE-POSITIVE OR PSEUDO-CALCIFICATION REFERS TO THOSE CASED IN WHICH OTHER PATHOLOGY IS MISTAKEN FOR CALCIFICATION OR FALSE-POSITIVE STAINING FOR CALCIUM. BASED ON THE INFORMATION PROVIDED, THE EXACT ROOT CAUSE FOR THIS EVENT COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HYDROVIEW INTRAOCULAR LENS WAS EXPLANTED DUE TO OPACIFICATION. THE LOT AND SERIAL NUMBER WERE NOT PROVIDED, THEREFORE, IT HAS NOT BEEN POSSIBLE TO DETERMINE WHETHER THE LENS MODEL IS HYDROVIEW 1.0. ADDITIONAL INFORMATION WAS REQUESTED, BUT HAS NOT BEEN RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
374989 HYDROVIEW INTRAOCULAR LENS INTRAOCULAR LENS HQL BAUSCH & LOMB H60M

Patients

Seq Age Sex Outcome Treatment
1 Other