ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-19540
- Event Type
- Malfunction
- Date Received
- June 12, 2015
- Report Date
- May 29, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2015 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE PUMP SHOWED THAT THE CONTRAST BUTTON SYMBOL WAS FADING; WHICH HAS NO EFFECT ON INSULIN DELIVERY. DURING TESTING, ALL THE KEYPAD BUTTONS RESPONDED PROPERLY. THE KEYPAD COVER WAS REMOVED AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. EVIDENCE OF MOISTURE WAS OBSERVED IN THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE OK, UP AND DOWN ARROW KEYPAD BUTTONS WERE UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 385657 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |