FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4839807 · Received June 12, 2015

Report

Report Number
2939301-2015-23946
Event Type
Malfunction
Date Received
June 12, 2015
Report Date
June 2, 2015
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER READ INACCURATELY ERRATIC. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿132, 133, 140 AND 203 MG/DL ¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR PRECISION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385871 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3748223

Patients

Seq Age Sex Outcome Treatment
1