FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PROFILE
MDR report key: 483512
·
Received September 8, 2003
Report
- Report Number
- 2939301-2003-06775
- Event Type
- Malfunction
- Date Received
- September 8, 2003
- Report Date
- September 5, 2003
- Manufacturer
- LIFESCAN, INC.
- Product Code
- CFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- PATIENT
Narratives
Description of Event or Problem · 1
A PT REPORTED BLOOD GLUCOSE RESULTS OF 187, 203, 187, 333, 141MG/DL AND 334MG/DL WITH A LIFESCAN METER, PERFORMED WITHIN 11-20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <=20% AND/OR <=20MG/DL, THEREBY INDICATING A POTENTIAL MALFUNCTION OF THE METER IN TERMS OF PRECISION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PROFILE | GLUCOSE MONITORING SYS/KIT | CFR | LIFESCAN, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |