FDA Adverse Event Injury Summary report: N

SELECTSECURE

MDR report key: 4834387 · Received June 10, 2015

Report

Report Number
2649622-2015-06699
Event Type
Injury
Date Received
June 10, 2015
Date of Event
April 1, 2015
Report Date
April 1, 2015
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P030036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. OUTER INSULATION CUT AT 35.8CM AND BLOOD VISIBLE IN PROXIMAL CONDUCTOR, CUT MAY HAVE OCCURRED DURING IMPLANT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD IMPEDANCE DECREASED SIGNIFICANTLY. THE LEAD WAS EXPLANTED AND REPLACED. IT WAS NOTED THERE BLOOD IN THE LEAD SEEN UPON LEAD REPLACEMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376302 SELECTSECURE DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 383069

Patients

Seq Age Sex Outcome Treatment
1 00063 YR Hospitalization| R