FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4834245 · Received June 10, 2015

Report

Report Number
2649622-2015-06752
Event Type
Injury
Date Received
June 10, 2015
Date of Event
April 16, 2015
Report Date
April 16, 2015
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD SENSING DECREASED. THERE WAS ALSO REPORT OF UNDERSENSING AND OVERSENSING WITH MYOGENIC POTENTIAL. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375134 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00088 YR Hospitalization| R E2SR01 IPG