BD LOGIC BLOOD GLUCOSE MONITOR

MDR report key: 483413 · Received September 9, 2003

Report

Report Number
2243072-2003-00050
Event Type
Other
Date Received
September 9, 2003
Date of Event
August 8, 2003
Report Date
September 5, 2003
Manufacturer
BECTON DICKINSON
Product Code
CFR
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER WAS FEELING THE BLOOD READINGS WERE WRONG, RAN COMPARISON AGAINST COMPETITIVE METER. ANALYSIS RUN ON CLARKE ERROR USING REPORTED COMPETITIVE READINGS (56) AS REF. POINT VS. BD READING (110, 118) YIELDED DATA POINTS IN "D" RANGE OF THE GRID, OUTSIDE ACCEPTABLE RANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC BLOOD GLUCOSE MONITOR BD TEST STRIPS CFR BECTON DICKINSON NA 1078360

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other