FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 4833967 · Received June 10, 2015

Report

Report Number
2649622-2015-06873
Event Type
Malfunction
Date Received
June 10, 2015
Date of Event
April 6, 2015
Report Date
April 6, 2015
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INITIAL IMPLANT THE HELIX ON THE RIGHT VENTRICULAR (RV) LEAD WOULD NOT EXTEND. THE RV LEAD WAS NOT USED AND REPLACED WITH ANOTHER LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380019 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1 00066 YR