FDA Adverse Event Injury Summary report: N

CAPSURE SP NOVUS

MDR report key: 4832958 · Received June 10, 2015

Report

Report Number
2649622-2015-07321
Event Type
Injury
Date Received
June 10, 2015
Date of Event
April 8, 2015
Report Date
April 9, 2015
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE AT THE FIRST RIB/CLAVICLE LOCATION WHILE IN VIVO. THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE AT THE FIRST RIB/CLAVICLE LOCATION WHILE IN VIVO, AND THE INNER INSULATION OF THE LEAD DEVELOPED A BREACH AT THE FIRST RIB/CLAVICLE LOCATION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH AT THE FIRST RIB/CLAVICLE LOCATION WHILE IN VIVO, AND THE OUTER INSULATION OF THE LEAD WAS COSMETICALLY IMPACTED AT THE FIRST RIB/CLAVICLE SITE WHILE IN VIVO. THE INNER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION, AND THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION. ANALYST COMMENTED, FRACTURE WAS CONFIRMED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH, VARYING IMPEDANCE, HIGH SHORT INTERVAL COUNTS (SIC) DUE TO SUB-CLAVIAN CRUSH, AND FRACTURE. IT WAS ALSO REPORTED THAT THE ATRIAL LEAD EXHIBITED NOISE, HIGH, VARYING IMPEDANCE AND FRACTURE. THE RV AND ATRIAL LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380681 CAPSURE SP NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 559453

Patients

Seq Age Sex Outcome Treatment
1 00060 YR Hospitalization| R (B)(4) ICD, 694765 LEAD