FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 4832735
·
Received June 10, 2015
Report
- Report Number
- 2182208-2015-02106
- Event Type
- Malfunction
- Date Received
- June 10, 2015
- Date of Event
- January 23, 2015
- Report Date
- March 6, 2015
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: DURING ANALYSIS IT WAS CONFIRMED THAT THE HEAD WAS ONLY INTERMITTENTLY ABLE TO INTERROGATE DEVICES AND IT FAILED ITS UPLINK TESTS. THE CABLE WAS REPLACED TO RESOLVE. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RADIOFREQUENCY PROGRAMMER HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376387 | CARELINK | PROGRAMMER, PACEMAKER | KRG | MEDTRONIC, INC. | 2067L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2090AA PROGRAMMER |