FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 4832735 · Received June 10, 2015

Report

Report Number
2182208-2015-02106
Event Type
Malfunction
Date Received
June 10, 2015
Date of Event
January 23, 2015
Report Date
March 6, 2015
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: DURING ANALYSIS IT WAS CONFIRMED THAT THE HEAD WAS ONLY INTERMITTENTLY ABLE TO INTERROGATE DEVICES AND IT FAILED ITS UPLINK TESTS. THE CABLE WAS REPLACED TO RESOLVE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RADIOFREQUENCY PROGRAMMER HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376387 CARELINK PROGRAMMER, PACEMAKER KRG MEDTRONIC, INC. 2067L

Patients

Seq Age Sex Outcome Treatment
1 2090AA PROGRAMMER