FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4832472 · Received June 10, 2015

Report

Report Number
2939301-2015-23539
Event Type
Malfunction
Date Received
June 10, 2015
Report Date
May 28, 2015
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE LAY USER/PATIENTS METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2. THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE (UNKNOWN BRAND). THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿312 MG/DL¿ WITH THE SUBJECT METER AND ¿132 MG/DL¿ ON THE OTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380210 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1