PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2182208-2015-01615
- Event Type
- Malfunction
- Date Received
- June 10, 2015
- Date of Event
- March 30, 2015
- Report Date
- April 22, 2015
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-1661_2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: DURING ANALYSIS THE DEVICE FAILED AN INCOMING VXI TEST DUE TO INTERMITTENT OPERATION. ANALYSIS ALSO NOTED THAT THE UPPER CASE WAS DENTED, BROKEN AND CONTAMINATED, THE LOWER CASE AND BATTERY DRAWER WERE BROKEN AND CONTAMINATED, THE HEARTWIRE BLOCK WAS CONTAMINATED AND THE SERIAL NUMBER LABEL WAS TORN. THE DEVICE WAS TO BE SCRAPPED IN-HOUSE. (B)(4).
THE EXTERNAL PULSE GENERATOR WAS RETURNED AS A TRADE-IN AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377315 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC, INC. | 5348 |