ATTAIN ABILITY
Report
- Report Number
- 2649622-2015-05436
- Event Type
- Injury
- Date Received
- June 10, 2015
- Date of Event
- April 29, 2015
- Report Date
- April 29, 2015
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING AND APPARENT EXPLANT DAMAGE.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IT WAS REPORTED THAT ONE DAY POST IMPLANT IS WAS FOUND THAT THE LEFT VENTRICULAR (LV) LEAD HAD NO CAPTURE AT MAXIMUM OUTPUT AND CHEST X-RAY SHOWED THAT THE LEAD HAD PULLED BACK APPROXIMATELY 1-2 CENTIMETERS. THE LEAD WAS REMOVED AND ANOTHER LEAD WAS ATTEMPTED, HOWEVER IT MOVED AFTER PLACEMENT. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 374654 | ATTAIN ABILITY | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 419688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00046 YR | Hospitalization| R | DTBA1D4 ICD, 6947M62 LEAD |