FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 4831553 · Received June 10, 2015

Report

Report Number
2649622-2015-05436
Event Type
Injury
Date Received
June 10, 2015
Date of Event
April 29, 2015
Report Date
April 29, 2015
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING AND APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT ONE DAY POST IMPLANT IS WAS FOUND THAT THE LEFT VENTRICULAR (LV) LEAD HAD NO CAPTURE AT MAXIMUM OUTPUT AND CHEST X-RAY SHOWED THAT THE LEAD HAD PULLED BACK APPROXIMATELY 1-2 CENTIMETERS. THE LEAD WAS REMOVED AND ANOTHER LEAD WAS ATTEMPTED, HOWEVER IT MOVED AFTER PLACEMENT. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
374654 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00046 YR Hospitalization| R DTBA1D4 ICD, 6947M62 LEAD