CARELINK
Report
- Report Number
- 2182208-2015-01677
- Event Type
- Malfunction
- Date Received
- June 10, 2015
- Date of Event
- March 9, 2015
- Report Date
- April 15, 2015
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE SYSTEM FAN WAS NOISY. ANALYSIS FOUND THE ECG (ELECTROCARDIOGRAM) CONNECTOR ON THE LEM (LINK ELECTRONIC MODULE) PRINTED CIRCUIT BOARD ASSEMBLY WAS LOOSE. (B)(4).
IT WAS REPORTED THE FAN IS NOISY. THE PROGRAMMER WAS RETURNED TO THE MANUFACTURER FOR REPAIR, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377613 | CARELINK | PROGRAMMER, PACEMAKER | KRG | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 229047 ANALYZER |