FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 4830840 · Received June 10, 2015

Report

Report Number
2649622-2015-05860
Event Type
Injury
Date Received
June 10, 2015
Date of Event
April 3, 2015
Report Date
April 6, 2015
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS REVEALED THE RIGHT VENTRICULAR LEAD INTEGRITY ALERT WAS TRIGGERED, THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE, THE PACING CAPTURE THRESHOLD IN THE RV (RIGHT VENTRICLE) WAS ELEVATED, AND THERE WAS OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). (B)(4)

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: D224DRG, ICD, IMPLANTED: (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD HAD AN APPARENT FRACTURE WITH NO PACING CAPTURE AND OVERSENSING NOTED. THE LEAD WAS CAPPED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, THE RV LEAD HELIX WOULD NOT EXTEND AND HIGH IMPEDANCE WAS OBSERVED. ANOTHER RV LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373854 SPRINT DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6945-65

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R