BD LOGIC (COBRANDED) BLOOD GLUCOSE METER
MDR report key: 482665
·
Received September 3, 2003
Report
- Report Number
- 2243072-2003-00042
- Event Type
- Other
- Date Received
- September 3, 2003
- Date of Event
- July 30, 2003
- Report Date
- August 29, 2003
- Manufacturer
- BECTON DICKINSON
- Product Code
- CFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER QUESTIONING ACCURACY OF THE METER. ANALYSIS RUN ON CLARKE ERROR USING REPORTED LAB READINGS (57, 62, 42) AS REF. POINT VS. BD READING (200, 225, 290) YIELDED DATA POINTS IN "E" ZONE OF THE GIRD, OUTSIDE ACCEPTABLE RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC (COBRANDED) BLOOD GLUCOSE METER | BD TEST STRIPS | CFR | BECTON DICKINSON | NA | 3078041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |