BABY CONTROL COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2015-00274
- Event Type
- Malfunction
- Date Received
- June 8, 2015
- Report Date
- May 11, 2015
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DATE OF REPORT WAS CHANGED FROM (B)(6) 2015 TO (B)(6) 2015. THE PRODUCT DESCRIPTION WAS CHANGED FROM MANUAL CONTROLLED COSYCOT INFANT WARMER TO BABY CONTROL COSYCOT INFANT WARMER. THE PRODUCT MODEL AND CATALOGUE NUMBER WERE CHANGED TO FROM IW950AEK TO IW932AEK. METHOD: THE COMPLAINT IW932AEK BABY CONTROL COSYCOT INFANT WARMER WAS SERVICED BY A QUALIFIED SERVICE ENGINEER OF FPH UK AT THE HOSPITAL'S FACILITY. OUR ANALYSIS IS BASED ON THE SERVICE REPORT AND PHOTOGRAPHS PROVIDED BY FPH SERVICE ENGINEER, AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS. RESULTS: PHOTOGRAPHS PROVIDED BY FPH SERVICE ENGINEER REVEALED THAT THERE WERE LONGITUDINAL CRACKS AT THE CENTER OF THE COLUMN OF THE PLASTIC LEGS. THE REVIEW OF OUR PRODUCTION RECORD SHOWED THAT THERE WAS NO DISCREPANCY WITH THE SUBJECT INFANT WARMER WHEN IT WAS RELEASED FOR DISTRIBUTION MORE THAN 10 YEARS AGO. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 041116. CONCLUSION: CRACKS TO THE PLASTIC LEG BASE REGION OF THE COSYCOT INFANT WARMER ARE KNOWN TO OCCUR WHEN IT HAS BEEN OVERLOADED. THE MAXIMUM WEIGHT LIMIT FOR THE INFANT WARMERS IS SPECIFIED IN THE INFANT WARMER PRODUCT TECHNICAL MANUAL (PTM). NO PATIENT CONSEQUENCE WAS REPORTED AS A RESULT OF THIS INCIDENT. THE INFANT WARMER WAS RETURNED INTO HOSPITAL SERVICE AFTER THE PLASTIC BASE WAS REPLACED WITH A METAL BASE AND THE UNIT PASSED THE PERFORMANCE CHECKS SPECIFIED IN THE PTM. SERVICING CONDUCTED AT THE HOSPITAL.
(B)(4). THE COMPLAINT IW932AEK MANUAL CONTROLLED COSYCOT INFANT WARMER HAS ONLY RECENTLY BEEN SERVICED BY A QUALIFIED SERVICE ENGINEER FROM FISHER & PAYKEL HEALTHCARE SERVICE CENTRE IN THE (B)(6). WE ARE CURRENTLY IN THE PROCESS OF OBTAINING FURTHER INFORMATION IN ORDER TO ASSIST US WITH THE ANALYSIS AND IDENTIFICATION OF A POSSIBLE ROOT CAUSE OF THE REPORTED EVENT. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.
A HOSPITAL IN THE (B)(6) REPORTED THAT THE REAR LEGS OF AN IW932AEK MANUAL CONTROLLED COSYCOT INFANT WARMER WERE CRACKED. NO PATIENT CONSEQUENCE WAS REPORTED. THE HOSPITAL ALSO REQUESTED TO REPAIR THE UNIT.
A HOSPITAL IN THE (B)(6) REPORTED THAT THE REAR LEGS OF AN IW932AEK BABY CONTROL COSYCOT INFANT WARMER WERE CRACKED. NO PATIENT CONSEQUENCE WAS REPORTED. THE HOSPITAL ALSO REQUESTED TO REPAIR THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 367974 | BABY CONTROL COSYCOT INFANT WARMER | FMT | FMT | FISHER & PAYKEL HEALTHCARE LIMITED | IW932AEK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |