FDA Adverse Event Injury Summary report: N

GALT GUIDEWIRE 0.018" X 60 CM

MDR report key: 4824996 · Received May 29, 2015

Report

Report Number
MW5042953
Event Type
Injury
Date Received
May 29, 2015
Date of Event
May 4, 2015
Report Date
May 29, 2015
Manufacturer
GALT
Product Code
DQX
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PICC PLACED TO LEFT LOWER EXTREMITY, PLACEMENT WAS EASY. UPON ATTEMPT TO PULL THE WIRE IT WAS DIFFICULT TO PULL, SO PICC LINE WAS ADJUSTED. SECOND ATTEMPT TO PULL WIRE, MORE BRISKLY THE WIRE PULLED OUT. UPON INSPECTION OF THE WIRE, IT WAS FRAYED AND WAS NOT 60CM IN LENGTH. DR( B)(6) WAS MADE AWARE. XRAYS ORDERED. WIRE ON CHEST XRAY APPEARS TO BE IN-SITU IN HEART. INTERVENTIONAL RADIOLOGY AND CARDIOLOGY CONSULTED BY PICU. WIRE UTILIZED WAS A GALT GUIDEWIRE. CARDIAC CATH LAB UTILIZATION TO RETRIEVE BROKEN WIRE (B)(6) 2015 22:50 TO (B)(6) 2015 AT 01:08. GUIDEWIRE STILL BEING UTILIZED FOR INSERTION OF PICC LINES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
351016 GALT GUIDEWIRE 0.018" X 60 CM GALT GUIDEWIRE 0.018" X 60 CM DQX GALT SGW-019-32

Patients

Seq Age Sex Outcome Treatment
1 4 YR Required Intervention