TRUERESULT
Report
- Report Number
- 1052693-2015-00834
- Event Type
- Malfunction
- Date Received
- June 4, 2015
- Date of Event
- May 5, 2015
- Report Date
- December 16, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT NOT YET RETURNED FOR EVAL. (B)(4).
(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAD AN INACCURATE REFERENCE OR USER REUSED TEST STRIP OR USER'S TEST STRIP HAD POOR FILL. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 115-150MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 11/30/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST. 91MG/DL AND 93MG/DL FASTING. REVIEWED METER MEMORY: 1: 65MG/DL, (B)(6) 2015, 11:41:55 AM, FASTING: NO; 2: 63MG/DL, (B)(6) 2015, 11:41:55 AM, FASTING: NO; 3: 105MG/DL, (B)(6) 2015, 11:41:55 AM, FASTING: YES; 4: 91MG/DL, (B)(6) 2015, 11:41:55 AM, FASTING: YES; 5: 109MG/DL, (B)(6) 2015, 11:41:55 AM, FASTING: YES.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 115-150MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 11/30/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED ON (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 91MG/DL AND 93MG/DL FASTING. REVIEWED METER MEMORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 363890 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PR2183 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |