FDA Adverse Event Malfunction Summary report: N

TRUEBALANCE

MDR report key: 4824660 · Received June 4, 2015

Report

Report Number
1052693-2015-00850
Event Type
Malfunction
Date Received
June 4, 2015
Date of Event
May 8, 2015
Report Date
December 17, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K090495
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). NO PRODUCT YET RETURNED.

Additional Manufacturer Narrative · 1

(B)(4) MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. CUSTOMER'S CAREGIVER STATES THAT THE PATIENT FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S' EXPECTED BLOOD RESULTS ARE 78-120 MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 10/31/2016. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED 1 MONTH AGO. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 457 MG/DL AND 471 MG/DL FASTING. REVIEWED METER MEMORY: 1:322MG/DL, (B)(6) 2015, 04:39:00 AM FASTING: YES. 2:571MG/DL, (B)(6) 2015, 04:40:00 PM FASTING: YES. 3:422MG/DL, (B)(6) 2015, 03:08:00 PM FASTING: YES. 4:243MG/DL, (B)(6) 2015, 04:19:00 AM FASTING: YES. 5:429MG/DL, (B)(6) 2015, 04:52:00 PM FASTING: YES. NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER'S CAREGIVER STATES THAT THE PATIENT FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 78-120MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 10/31/2016. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED 1 MONTH AGO. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 457MG/DL AND 471MG/DL FASTING. REVIEWED METER MEMORY: 1:322MG/DLMG/DL (B)(6) 2015 04:39:00 AM FASTING:YES. 2:571MG/DLMG/DL (B)(6) 2015 04:40:00 PM FASTING:YES. 3:422MG/DLMG/DL (B)(6) 2015 03:08:00 PM FASTING:YES. 4:243MG/DLMG/DL (B)(6) 2015 04:19:00 AM FASTING:YES. 5:429MG/DLMG/DL (B)(6) 2015 04:52:00 PM FASTING:YES. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
363616 TRUEBALANCE BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUBALANCE BR4552

Patients

Seq Age Sex Outcome Treatment
1 0 YR