TRUEBALANCE
Report
- Report Number
- 1052693-2015-00850
- Event Type
- Malfunction
- Date Received
- June 4, 2015
- Date of Event
- May 8, 2015
- Report Date
- December 17, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K090495
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). NO PRODUCT YET RETURNED.
(B)(4) MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.
CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. CUSTOMER'S CAREGIVER STATES THAT THE PATIENT FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S' EXPECTED BLOOD RESULTS ARE 78-120 MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 10/31/2016. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED 1 MONTH AGO. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 457 MG/DL AND 471 MG/DL FASTING. REVIEWED METER MEMORY: 1:322MG/DL, (B)(6) 2015, 04:39:00 AM FASTING: YES. 2:571MG/DL, (B)(6) 2015, 04:40:00 PM FASTING: YES. 3:422MG/DL, (B)(6) 2015, 03:08:00 PM FASTING: YES. 4:243MG/DL, (B)(6) 2015, 04:19:00 AM FASTING: YES. 5:429MG/DL, (B)(6) 2015, 04:52:00 PM FASTING: YES. NO ADVERSE EVENT REPORTED.
CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER'S CAREGIVER STATES THAT THE PATIENT FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 78-120MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 10/31/2016. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED 1 MONTH AGO. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 457MG/DL AND 471MG/DL FASTING. REVIEWED METER MEMORY: 1:322MG/DLMG/DL (B)(6) 2015 04:39:00 AM FASTING:YES. 2:571MG/DLMG/DL (B)(6) 2015 04:40:00 PM FASTING:YES. 3:422MG/DLMG/DL (B)(6) 2015 03:08:00 PM FASTING:YES. 4:243MG/DLMG/DL (B)(6) 2015 04:19:00 AM FASTING:YES. 5:429MG/DLMG/DL (B)(6) 2015 04:52:00 PM FASTING:YES. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 363616 | TRUEBALANCE | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUBALANCE | BR4552 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |