FDA Adverse Event Malfunction Summary report: N

TRUETRACK

MDR report key: 4824601 · Received June 4, 2015

Report

Report Number
1052693-2015-00872
Event Type
Malfunction
Date Received
June 4, 2015
Date of Event
May 11, 2015
Report Date
January 5, 2016
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K032657
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 115 TO 159 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. COMPARING BLOOD GLUCOSE TEST RESULTS FROM TRUETRACK METER TO TRUERESULT METER. BACK TO BACK BLOOD TEST PERFORMED DURING CALL ON (B)(6) 2015 PRODUCED TEST RESULTS OF 310 MG/DL USING TRUETRACK METER AND 212 MG/DL USING TRUERESULT METER. VERIFIED STORAGE OF PRODUCT IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 07/31/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 1:310MG/DL (B)(6) 2015 11:39AM; 2:123MG/DL (B)(6) 2015 11:36AM; 3:129MG/DL (B)(6) 2015 11:32AM; 4:244MG/DL (B)(6) 2015 11:07AM; 5:118MG/DL (B)(6) 2015 11:04AM; ADVERSE EVENT NOT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 115 TO 159 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. COMPARING BLOOD GLUCOSE TEST RESULTS FROM TRUETRACK METER TO TRUERESULT METER. BACK TO BACK BLOOD TEST PERFORMED DURING CALL ON (B)(6) 2015 PRODUCED TEST RESULTS OF 310 MG/DL USING TRUETRACK METER AND 212 MG/DL USING TRUERESULT METER. VERIFIED STORAGE OF PRODUCT IS INSTRUCTED SPECIFICATION,.TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 07/31/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 310MG/DL, (B)(6) 2015, 11:39 AM; : 310MG/DL, (B)(6) 2015, 11:39 AM; 123MG/DL, (B)(6) 2015, 11:36 AM; 244MG/DL, (B)(6) 2015, 11:07 AM; 118MG/DL, (B)(6) 2015, 11:04 AM. ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
363654 TRUETRACK BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUETRACK RS4604

Patients

Seq Age Sex Outcome Treatment
1 0 YR