PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-63359
- Event Type
- Malfunction
- Date Received
- June 8, 2015
- Date of Event
- May 19, 2015
- Report Date
- May 19, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONDS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED DURING TESTING. THE DEVICE HAD MINOR SCRATCHES ON THE LCD WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNER, AND CRACKED RESERVOIR TUBE LIP.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER STATED THAT SHE HAD TAKEN THE INSULIN PUMP OFF TO GO SWIMMING, AND WHEN SHE ATTEMPTED TO PUT THE DEVICE BACK ON, THE BUTTON ERROR ISSUES BEGAN. THE CUSTOMER DID NOT KNOW THE BLOOD GLUCOSE READING. THE CUSTOMER DID NOT OBSERVE ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED TO REPLACE THE INSULIN PUMP AND AGREED TO RETURN THE DEVICE FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 368942 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |