FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4823188 · Received June 6, 2015

Report

Report Number
2531779-2015-18851
Event Type
Malfunction
Date Received
June 6, 2015
Report Date
May 27, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DISPLAY ISSUE) ISSUE. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT WITH THIS COMPLAINT. CUSTOMER SUPPORT HAS MADE SEVERAL ATTEMPTS TO CONTACT THE REPORTER IN FOLLOW UP, HOWEVER, THE REPORTER DID NOT RESPOND. NO FURTHER INFORMATION WAS AVAILABLE; IF FURTHER INFORMATION IS PROVIDED A FOLLOW UP REPORT SHALL BE MADE. THIS COMPLAINT IS BEING REPORTED BECAUSE OF THE ALLEGATION OF A DISPLAY ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366835 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 51 YR