PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4823153 · Received June 6, 2015

Report

Report Number
2032227-2015-18101
Event Type
Injury
Date Received
June 6, 2015
Date of Event
May 19, 2015
Report Date
May 19, 2015
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD MOISTURE DAMAGE ON KEYPAD TRACES. ALL BUTTONS WERE RESPONDING PROPERLY. THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE ON THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE HAD A KEYPAD ANOMALY ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE WAS 500 MG/DL. CUSTOMER STATED THAT SHE HAD A NO DELIVERY ALARM AND SHE COULD NOT DO ANYTHING WITH THE PUMP. CUSTOMER IS NOT ABLE TO SEE ANYTHING AND THE BUTTONS ARE NOT WORKING, ESPECIALLY THE ACT AND ESC BUTTONS. CUSTOMER WAS ADVISED THAT THE PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ALSO ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366912 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAS

Patients

Seq Age Sex Outcome Treatment
1 31 YR