FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4823144 · Received June 6, 2015

Report

Report Number
2531779-2015-18874
Event Type
Malfunction
Date Received
June 6, 2015
Report Date
May 27, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 09/25/2015 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: THE REPORTED EVENT DATE OF (B)(6) 2015 WAS FOUND TO BE OVERWRITTEN IN THE BLACK BOX. BLACK BOX INDICATED MULTIPLE ¿CALL SERVICE (CS) 162¿ LOSS OF PRIME WARNINGS DUE TO LOW NON ZERO FORCE. THE RETURNED BATTERY/CARTRIDGE CAPS WERE USED DURING INVESTIGATION. THE ¿EZ-PRIME¿ STEPS WERE PERFORMED CORRECTLY ON THE PUMP. THERE WERE NO ¿CS162¿ WARNINGS OR ANY ERRORS, ALARMS OR WARNINGS THAT OCCURRED DURING THE INVESTIGATION. THE PRODUCT PERFORMED WITHIN SPECIFICATION AND THE REPORTED ¿LOSS OF PRIME¿ COMPLAINT WAS UNABLE TO BE DUPLICATED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THE REPORTER ALLEGED THAT APPROXIMATELY 1 UNIT OF INSULIN WAS INFUSED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366909 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 84 YR