FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4823110 · Received June 6, 2015

Report

Report Number
3004209178-2015-63326
Event Type
Malfunction
Date Received
June 6, 2015
Date of Event
May 19, 2015
Report Date
May 19, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. THE INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND, AND BASIC OCCLUSION TESTS, NO ALARMS NOTED DURING TESTING. ALL OF THE BUTTONS FUNCTIONED PROPERLY AND NO MOISTURE DAMAGE WAS NOTED ON THE KEYPAD TRACES, INSIDE RESERVOIR COMPARTMENT, OR MOTOR NOTED. THE DEVICE HAD MINOR SCRATCHES ON THE LCD WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THE INSULIN PUMP HAD A MOISTURE DAMAGE. CUSTOMER'S BLOOD GLUCOSE WAS 286 MG/DL. THE CUSTOMER STATED THAT THE INSULIN PUMP WAS EXPOSED TO INSULIN . CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACKUP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366779 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523LNAS

Patients

Seq Age Sex Outcome Treatment
1 79 YR