PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-18123
- Event Type
- Malfunction
- Date Received
- June 6, 2015
- Date of Event
- May 19, 2015
- Report Date
- May 19, 2015
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED DURING TESTING. THE PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE LCD WINDOW, A CRACKED LCD WINDOW, AND A SCRATCHED RESERVOIR TUBE WINDOW.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER EXPLAINED THAT THE INSULIN PUMP WAS SCROLLING CONSTANTLY AND THEN ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 123 MG/DL. WHILE TROUBLESHOOTING, THE CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER HEALTHCARE PROVIDER'S INSTRUCTIONS. THE CUSTOMER WAS INFORMED THE DEVICE WOULD BE REPLACED. THE CUSTOMER AGREED TO RETURN THE INSULIN PUMP FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 366952 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723RNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |