FDA Adverse Event Injury Summary report: N

HOMECHOICE INTEGRATED APD SET WITH CASSETTE 3-PRONG

MDR report key: 482240 · Received September 3, 2003

Report

Report Number
1423500-2003-00939
Event Type
Injury
Date Received
September 3, 2003
Date of Event
August 5, 2003
Report Date
August 6, 2003
Manufacturer
BAXTER HEALTHCARE CORP.
Product Code
FKX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

A HOME PATIENT (HP) CALLED BAXTER TECHNICAL SERVICE TO REPORT THAT THEY RECEIVED AIR IN THEIR DAY FILL 1 OF THEIR DAY FILL 1 OF APD THERAPY WITH THE HOMECHOICE MACHINE ON 2 CONSECTIVE DAYS. HP RELATES THAT PATIENT HAVE A DRY DAY AND PERFORMS A BRIEF INITIAL DRAIN ONLY PRIOR TO THERAPY ADVANCING INTO THEIR DAY FILL 1. HP HAS AN INITIAL DRAIN SET POINT AT 1500ML TO ENSURE SOME DRAINAGE PRIOR TO FILL, BUT BYPASSES THIS DRAIN IF NO FLUID DRAINS AFTER 1 MINUTE OR SO. HP STATES THEY ARE SURE THAT THE PATIENT LINE OF THEIR HOMECHOICE INTEGRATED SET WAS COMPLETELY PRIMED PRIOR TO FILL 1, BUT PATIENT SAW TWO AIR BUBBLES FROM THE PATIENT LINE GO INTO THEM ON BOTH OCCASIONS. HP DESCRIBES THE BUBBLES AS APPROXIMATELY 1/2 INCH IN LENGTH IN THE PATIENT LINE AND THEY WERE UNABLE TO STOP TREATMENT BEFORE THE AIR BUBBLES WERE INFUSED. HP REPORTS THAT PATIENT IMMEDIATELY FELT PAIN IN THEIR SHOULDER AND ARM AFTER THIS OCCURRED AND CONTINUED TO EXPERIENCE PAIN FOR THE FOLLOWING WEEK. APD THERAPY WAS COMPLETED WITH THE INITIAL SET USED IN BOTH INCIDENTS. HP REPORTS THAT ATTEMPTED TO PURGE THE AIR BY RIDING THEIR BIKE AND NO MEDICAL INTERVENTION WAS REQUIRED. HP SWITCHED THEIR SET TO A DIFFERENT LOT NUMBER AND REPORTED NO FURTHER DIFFICULTIES. HP REPORTS THEY NOTED NOTHING UNUSUAL WITH HOMECHOICE SETS USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE INTEGRATED APD SET WITH CASSETTE 3-PRONG HOMECHOICE CASSETTE FKX BAXTER HEALTHCARE CORP. NA H03D16021

Patients

Seq Age Sex Outcome Treatment
1 78 YR Other DIANEAL LOW CALCIUM SOLUTION (STRENGTH UNK) 2003,| MINICAP DISCONNECT CAP, 2003.| HOMECHOICE AUTOMATED PD SYSTEM 115 VOLT, 2003,