530G INSULIN PUMP
Report
- Report Number
- 3004209178-2015-63034
- Event Type
- Malfunction
- Date Received
- June 5, 2015
- Date of Event
- May 17, 2015
- Report Date
- May 18, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE FOUND ON THE ELECTRONICS, MOTOR, BATTERY TUBE, AND VIBRATOR. THE DEVICE HAD CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED BELT CLIP SLOT, AND CRACKED RESERVOIR TUBE LIP.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP EXPERIENCED KEYPAD ISSUES. THE CUSTOMER STATED THE ACT BUTTON WAS WORKING INTERMITTENTLY AND THAT THE UP AND DOWN BUTTONS WERE NOT WORKING. THE CUSTOMER EXPLAINED THAT THE INSULIN PUMP WAS SCROLLING THROUGH EVERYTHING. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. WHILE TROUBLESHOOTING, THE CUSTOMER EXPLAINED THAT HE WAS RUNNING A MARATHON AND THAT SWEAT MAY HAVE COME INSIDE THE SCREEN. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER HEALTHCARE PROVIDER'S INSTRUCTIONS. THE CUSTOMER WAS INFORMED THE DEVICE WOULD BE REPLACED. THE CUSTOMER AGREED TO RETURN THE INSULIN PUMP FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 365664 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |