PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-63001
- Event Type
- Malfunction
- Date Received
- June 5, 2015
- Date of Event
- May 18, 2015
- Report Date
- May 18, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP RECEIVED WITH AN INTERMITTENT BUTTON DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE ON DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED VIA PHONE CALL, THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME WAS UNKNOWN. TROUBLESHOOTING OCCURRED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP, AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 364812 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |