FDA Adverse Event Injury Summary report: N

SURGITRON FFPF EMC

MDR report key: 4821949 · Received June 5, 2015

Report

Report Number
2428235-2015-00003
Event Type
Injury
Date Received
June 5, 2015
Date of Event
May 7, 2015
Report Date
June 4, 2015
Manufacturer
CYNOSURE, INC. DBA ELLMAN
Product Code
GEI
PMA / PMN Number
K972072
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CANINE UNDERWENT RIGHT FOREQUARTER AMPUTATION WITH THE EMC ELECTROSURGICAL GENERATOR AND THE NEUTRAL PAD WAS PLACED ON THE DOG'S BACK WITH A SMALL PORTION (APPROXIMATELY 2 INCH TRIANGLE) CONTACTING SHAVED SKIN WHILE THE REMAINDER CONTACTED FUR. THE BURN WAS LOCATED IN THE SHAVED AREA WHERE THE NEUTRAL PAD CONTACTED THE SKIN DIRECTLY, WAS MINOR IN NATURE AND DID NOT REQUIRE SUBSEQUENT INTERVENTIONS OR TREATMENT. THE NEUTRAL PAD WAS MISAPPLIED TO THE PATIENT (DOG) BY BEING PLACED OVER BOTH SHAVED AND FUR COVERED AREAS. THE DIFFERENCE IN IMPEDANCE OVER THE PAD CAUSED A CONCENTRATION OF ENERGY IN THE LOW IMPEDANCE AREA AND CAUSED A BURN. ALTHOUGH THIS EMC WAS BEING USED FOR VETERINARY APPLICATIONS, THE PRODUCT IS ALSO AVAILABLE FOR HUMAN APPLICATIONS. THE LIKELIHOOD OF A SIMILAR EVENT OCCURING IN A HUMAN IS LOW AS THERE IS LESS OPPORTUNITY FOR THE DRASTIC DIFFERENCE IN IMPEDANCES AS THAT BETWEEN FUR AND SKIN IN VETERINARY APPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
365037 SURGITRON FFPF EMC ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI CYNOSURE, INC. DBA ELLMAN IEC-NPC

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention