FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4821940 · Received June 5, 2015

Report

Report Number
2531779-2015-18775
Event Type
Injury
Date Received
June 5, 2015
Date of Event
May 27, 2015
Report Date
May 27, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 09/15/2015: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING RESULTS: BATTERY COMPARTMENT CRACK FOUND BELOW GRIP PAD TO THE CASE SEAL. UNABLE TO DUPLICATE THE COMPLAINT. THE TDD¿S ADD UP CORRECTLY AND REFLECT THE USERS PROGRAMMED BASAL RATES. PUMP PASSED DELIVERY ACCURACY TEST AND WAS FOUND TO BE DELIVERING ACCURATELY AND WITHIN RANGE. THE LAST BASAL DELIVERY AND THE LAST BOLUS DELIVERY WERE ON (B)(6) 2015. PUMP BOOTS TO THE VERIFY SCREEN WITH NO ISSUES. A REWIND, LOAD AND PRIME SEQUENCE WAS PERFORMED SUCCESSFULLY WITH NO ERRORS OR ALARMS. THE PUMP CLEARLY DISPLAYS MESSAGE ¿DISCONNECT INFUSION SET FROM YOUR BODY!" ON THE REWIND SCREEN AND ¿BE SURE SET IS DISCONNECTED FROM YOUR BODY. THEN SELECT CONTINUE" ON THE PRIME SCREEN. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT HAD A BLOOD GLUCOSE OF 40 MG/DL, NO SYMPTOMS, AFTER AN INADVERTENT INFUSION THAT DAY. THERE IS NO ALLEGATION OF PUMP MALFUNCTION. THIS COMPLAINT IS BEING REPORTED AS THE PATIENT EXPERIENCED HYPOGLYCEMIA SUBSEQUENT TO THE USE ERROR OF AN INADVERTENT INFUSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
365018 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 6 YR Life Threatening