FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION S7 SYSTEM

MDR report key: 4821939 · Received June 5, 2015

Report

Report Number
1723170-2015-00675
Event Type
Malfunction
Date Received
June 5, 2015
Date of Event
April 24, 2015
Report Date
February 22, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Product Code
HAW
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CORRECTION: IT WAS DISCOVERED ON 27-JAN-2016, THAT AN INCORRECT DATE WAS REFERENCED IN A PREVIOUS SUPPLEMENTAL 3500A SUBMISSION. THE REPORTED DATE OF 16-AUG-2016 WAS REPORTED INCORRECTLY AND SHOULD BE 16-AUG-2015. STATEMENT IN PREVIOUS SUPPLEMENTAL 3500A SUBMISSION SHOULD READ: ¿ON 30-JUN-2015, IT WAS NOTICED THAT A CODING ERROR IN MDR SUBMISSIONS FROM OUR FACILITY RESULTED IN THE MAUDE DATABASE INCORRECTLY CODING THE DEVICES RELATED TO OUR MDR SUBMISSIONS FROM 25-MAY-2015 TO 16-AUG-2015. THE DECISION TO WAIT UNTIL THE DATABASE WAS CORRECTED WAS MADE AFTER CONSULTATION WITH THE FDA AS ADVISED BY A CONSUMER SAFETY OFFICER WITH THE INFORMATION ANALYSIS BRANCH, DIVISION OF POST MARKET SURVEILLANCE, OFFICE OF SURVEILLANCE AND BIOMETRICS. AN IT SOLUTION WAS IMPLEMENTED ON 16-AUG-2015. THIS MDR WAS SUBMITTED TO CORRECT THE CODING ERROR. THERE IS NO NEW INFORMATION TO CHANGE THE PATIENT INFORMATION, EVENT DESCRIPTION AND/OR MANUFACTURER NARRATIVE THAT WAS PREVIOUSLY REPORTED.¿

Additional Manufacturer Narrative · 1

NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. A MEDTRONIC REPRESENTATIVE, FOLLOWING UP WITH THE SITE, REPORTED THAT THE POWER CORD HAD A CUT IN IT AND THE COPPER WIRE WAS EXPOSED. REPLACEMENT POWER CORD SHIPPED TO SITE ON (B)(4) 2015. NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR ANALYSIS AS THE PART WAS DISCARDED ON-SITE. A MEDTRONIC REPRESENTATIVE REPLACED THE POWER CORD AND PERFORMED A NAVIGATION SYSTEM CHECK-OUT, ALL AREAS PASSED. THE SYSTEM WAS REPORTED TO BE FULLY FUNCTIONAL AFTER THE REPLACEMENT OF THE CORD.

Description of Event or Problem · 1

A SITE REPRESENTATIVE REPORTED THAT THE NAVIGATION SYSTEM'S POWER CABLE WAS FRAYED. NO FURTHER DETAILS REGARDING THE EXTENT OF THE DAMAGE, OR HOW IT OCCURRED, WERE PROVIDED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
365537 STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. (LOUISVILLE) S7

Patients

Seq Age Sex Outcome Treatment
1