FDA Adverse Event Injury Summary report: N

ANIMAS VIBE

MDR report key: 4821352 · Received June 5, 2015

Report

Report Number
2531779-2015-18723
Event Type
Injury
Date Received
June 5, 2015
Date of Event
June 3, 2015
Report Date
June 4, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
P130007
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 ORIGINALLY SUBMITTED ON (B)(6) 2015, RESUBMITTED PER ESG REQUEST ON (B)(6) 2015. DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2015 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE TOTAL DAILY DOSE HISTORY SHOWED THAT THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER PROGRAMMED BASAL RATES. THE PUMP HISTORY SHOWED THAT ON (B)(6) 2015, THE PUMP WAS MANUALLY SUSPENDED AT 07:34 AND MANUALLY RESUMED AT 08:32. A DELIVERY ACCURACY TEST WAS PERFORMED AND THE PUMP WAS FOUND TO BE DELIVERING WITHIN SPECIFICATIONS. NO DELIVERY RELATED DEFECTS WERE FOUND DURING TESTING.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS INDICATING THAT THE PATIENT REQUIRED PHONE ADVICE FROM A HEALTH CARE PROFESSIONAL FOR AN ALLEGED A BLOOD GLUCOSE OF 26 MMOL/L WITH VOMITING, DROWSINESS / DIFFICULTY WAKING UP, INCREASED THIRST, AND LARGE KETONES RELATED TO INACCURATE DELIVERY. THE PUMP WAS REVIEWED WITH THE REPORTER WHO INDICATED THAT THE BASAL AND BOLUS AMOUNTS DID NOT MATCH WITH THE TOTAL DAILY DOSE HISTORY. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPERGLYCEMIA REQUIRING ADVICE FROM A HEALTH PROFESSIONAL RELATED TO AN INACCURATE DELIVERY ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
365012 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR Life Threatening