3.5MM LOCKING SCREW SLF-TPNG W/STARDRIVE RECESS 42MM
Report
- Report Number
- 2520274-2015-14267
- Event Type
- Injury
- Date Received
- June 5, 2015
- Report Date
- May 28, 2015
- Manufacturer
- SYNTHES USA
- Product Code
- JDS
- PMA / PMN Number
- PK072095
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORD REVIEW AND THE INVESTIGATION COULD NOT BE COMPLETED. AN INVESTIGATION SUMMARY WAS PERFORMED. THE INVESTIGATION OF THE COMPLAINT ARTICLES HAS SHOWN THAT: THE LCP PERIARTICULAR PROXIMAL HUMERUS PLATES ARE INDICATED FOR FRACTURES, FRACTURE DISLOCATIONS, OSTEOTOMIES, AND NONUNIONS OF THE PROXIMAL HUMERUS. THE TECHNIQUE GUIDE (J9083-B) RECOMMENDS INSERTING CORTEX SCREWS FIRST TO PULL THE PLATE TO THE BONE AND TO AID IN REDUCTION OF ANY DISTAL SHAFT FRAGMENTS. THE TECHNIQUE GUIDE ALSO STATES TO USE STANDARD AO SCREW INSERTION TECHNIQUE TO INSERT 3.5MM CORTEX SCREWS IN THE DCU PORTION OF THE COMBI HOLES. THE COMPLAINT DESCRIPTION STATES THAT THE SCREWS BROKE TEN MONTHS POST A HUMERAL FRACTURE REVISION. IT IS LIKELY THAT THE BONE DID NOT HEAL CAUSING THE 4 SCREWS DISTAL TO THE FRACTURE TO FATIGUE AND FAIL. NOT FOLLOWING THE RECOMMENDED TECHNIQUE MAY HAVE ALSO CONTRIBUTED TO THE SCREW BREAKAGE. (B)(4) WAS REVIEWED DURING THE EVALUATION AND NO ISSUES WERE FOUND. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT BROKEN SCREWS WERE DISCOVERED APPROXIMATELY TEN MONTHS POST A HUMERAL FRACTURE REVISION PROCEDURE OF (B)(6) 2014. THE PATIENT BEGAN TO EXPERIENCE PAIN (WITH NO TRAUMA OCCURRING) IN (B)(6) 2015; X-RAYS OF (B)(6) 2015 SHOWED FOUR BROKEN SCREWS. IT WAS REPORTED THAT THE CONSTRUCT CONSISTS OF A PLATE AND ELEVEN SCREWS. THE PATIENT SUSTAINED A RIGHT HUMERUS FRACTURE WHEN SHE TRIPPED AND FELL WHILE WALKING A DOG ON (B)(6) 2012. IT WAS REPORTED THAT ORIGINALLY IT WAS "LEFT TO HEAL ON ITS OWN". WHEN IT DID NOT HEAL, IT WAS DISCOVERED THAT THE "BICEPS WERE IMPALED BETWEEN THE BONE". IN 2013, A PROCEDURE INVOLVING A PLATE AND SCREWS WAS PERFORMED. AFTER ONE YEAR, THIS CONSTRUCT "FELL APART", AND A REVISION WAS PERFORMED IN (B)(6) 2014. THIS IS REPORT 6 OF 9 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 365051 | 3.5MM LOCKING SCREW SLF-TPNG W/STARDRIVE RECESS 42MM | NAIL,FIXATION,BONE | JDS | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |