FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4820272 · Received June 4, 2015

Report

Report Number
2531779-2015-18667
Event Type
Malfunction
Date Received
June 4, 2015
Report Date
May 25, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 09/03/2015 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/18/2015 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWED NO EVIDENCE OF A POWER EVENT. THE BATTERY CAP WAS ABLE TO FULLY TIGHTEN HOWEVER THE BATTERY CAP THREADS AND COIN SLOT WERE DAMAGED. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES DUPLICATED. UNRELATED TO THE INITIAL ALLEGATION, THE DISPLAY SCREEN WAS DIM AND DISCOLORED.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE PUMP HAD NO POWER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361297 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 21 YR