ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-18662
- Event Type
- Malfunction
- Date Received
- June 4, 2015
- Report Date
- May 20, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 08/04/2015-DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2015 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A REVIEW OF THE PUMP BLACK BOX DATA REVEALED OCCLUSIONS ASSOCIATED WITH HIGH FORCES. DURING TESTING, THE REWIND, LOAD, AND PRIME STEPS WERE COMPLETED SUCCESSFULLY WITH NO DUPLICATED OCCLUSION ALARMS. THE FORCE SENSOR CALIBRATION MEASURED LOW AND WAS NOT WITHIN SPECIFICATIONS. EVIDENCE OF THE COMPLAINT WAS FOUND IN THE BLACK BOX; HOWEVER, THE COMPLAINT WAS NOT DUPLICATED. UNRELATED TO THE COMPLAINT, THE PUMP EMITTED A CS 088 CALL SERVICE ALARM. EVIDENCE OF MOISTURE INGRESS WAS OBSERVED BEHIND THE DISPLAY LENS, AND THE CASING WAS CRACKED NEAR THE DISPLAY. A LEAK TEST WAS PERFORMED AND FAILED DUE TO A LEAK AT THE CRACK IN THE CASING. THE PUMP CASE WAS REMOVED AND DAMAGE DUE TO MOISTURE WAS FOUND ON THE PRINTED CIRCUIT BOARD.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A OCCLUSION (FREQUENT/PERSISTENT) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 362145 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |