FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4820240 · Received June 4, 2015

Report

Report Number
2531779-2015-18631
Event Type
Malfunction
Date Received
June 4, 2015
Report Date
May 21, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/21/2015 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. THE KEYPAD WAS DAMAGED BUT ALL OF THE KEYPAD BUTTONS WERE STILL RESPONSIVE. CONTAMINATION WAS FOUND UNDERNEATH THE CONTRAST BUTTON CONTACT. (B)(6)

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 05/21/2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
362140 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 79 YR