FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4820223 · Received June 4, 2015

Report

Report Number
3004209178-2015-62778
Event Type
Malfunction
Date Received
June 4, 2015
Date of Event
May 18, 2015
Report Date
May 18, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO MISSING SEGMENTS NOTED. HOWEVER, THE INSULIN PUMP RECEIVED WITH INTERMITTENT RESPONSE DUE TO MOISTURE KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH PUMP CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT THEIR SCREEN IS FADED. THE BLOOD GLUCOSE READING IS 97 MG/DL. THE INSULIN PUMP WILL BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361875 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 26 YR