FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4820222 · Received June 4, 2015

Report

Report Number
3004209178-2015-62777
Event Type
Malfunction
Date Received
June 4, 2015
Date of Event
May 17, 2015
Report Date
May 18, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM DURING BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. THE DEVICE HAD CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE LCD WINDOW, AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER STATES THAT THEY RECEIVED A LOW BATTERY ALARM. THE BLOOD GLUCOSE READING WAS 313 MG/DL. THE INSULIN PUMP WILL BE REPLACED.