FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4820221
·
Received June 4, 2015
Report
- Report Number
- 3004209178-2015-62776
- Event Type
- Malfunction
- Date Received
- June 4, 2015
- Date of Event
- May 7, 2015
- Report Date
- May 18, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH HIGH IDLE CURRENT AND UNEXPECTED OFF NO POWER ALARM DUE TO OPEN LCD DRIVER ON LCD BOARD. THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW. NO FAILED BATTERY TEST ALARM NOTED.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
CUSTOMER STATES THAT THEY RECEIVED A LOW BATTERY ALARM. THE BLOOD GLUCOSE READING WAS 313 MG/DL. THE INSULIN PUMP WILL BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 361606 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |