FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4820192 · Received June 4, 2015

Report

Report Number
3004209178-2015-62764
Event Type
Malfunction
Date Received
June 4, 2015
Date of Event
May 18, 2015
Report Date
May 18, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP ALSO WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER'S MOTHER REPORTED VIA PHONE CALL THAT THE CUSTOMER HAD A KEYPAD ANOMALY. THE MOTHER STATED THE BUTTONS ON THE INSULIN PUMP WAS UNRESPONSIVE. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. THE MOTHER REPORTED POSSIBLE MOISTURE EXPOSURE TO THE INSULIN PUMP. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED. CUSTOMER AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
362930 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 7 YR