FDA Adverse Event Malfunction Summary report: N

CS BED 9153650455

MDR report key: 4817471 · Received June 4, 2015

Report

Report Number
3003433498-2015-00075
Event Type
Malfunction
Date Received
June 4, 2015
Date of Event
November 3, 2014
Report Date
May 7, 2015
Manufacturer
INVACARE CONTINUING CARE
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP WILL BE SENT IF THE PRODUCT OR ADDITIONAL INFORMATION IS OBTAINED.

Description of Event or Problem · 1

AS A RESULT OF PIN HOLDING POWER HEAD ACTUATOR MECHANISM BREAKING OR POPPING OUT, END USER ALLEGEDLY DROPPED INTO A HORIZONTAL POSITION WHILE RAISING HEAD OF BED. HE APPARENTLY HAD SOME INITIAL PAIN AND SAW DOCTOR. THERE WAS NO REHAB OR MEDICATION PRESCRIBED. END USERS ISSUE AT THIS POINT IS NOISE THAT BED MAKES WHEN HE INITIATES RAISING HEAD OF BED. HE ALSO CLAIMS THAT THERE HAS BEEN OIL AND METAL SHAVINGS LEFT ON CARPET UNDER BED. DEALER TECHNICIAN REPLACED ORIGINAL PIN FOR ACTUATOR WITH A STRONGER MORE ROBUST BOLT AND ALSO REPLACED HEADBOARD GUIDE WHEELS WITH SUPERIOR URETHANE ONES TO HELP ELIMINATE NOISE. DEALER TECHNICIAN SAYS HE CANNOT FID SOURCE OF LEAKING OIL OR METAL SHAVINGS. DEALER TECHNICIAN CLAIMS THAT REASON BED IS MAKING A NOISE, IS A RESULT OF END USER'S WEIGHT BEING CLOSE TO OR OVER THE MAXIMUM WEIGHT CAPACITY OF BED WHEN ADDING WEIGHT OF THE SIDE RAILS AND MATTRESS. DEALER TECHNICIAN ALSO CLAIMS THAT BED MAKES NO NOISE WHEN LIFTING HIS OWN PERSONAL WEIGHT. ((B)(6))

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
361519 CS BED 9153650455 BED, AC-POWERED ADJUSTABLE HOSPITAL FNL INVACARE CONTINUING CARE HCBED

Patients

Seq Age Sex Outcome Treatment
1 61 Other