FDA Adverse Event
Malfunction
Summary report: N
ACQSIM
MDR report key: 481461
·
Received August 28, 2003
Report
- Report Number
- 1525965-2003-00010
- Event Type
- Malfunction
- Date Received
- August 28, 2003
- Date of Event
- July 31, 2003
- Report Date
- July 31, 2003
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- KPQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
ACQSIM-CT OPERATED WITH PREFERENCE OF AUTO SEND ON, TO SEND IMAGES TO VOXELQ. CUSTOMER IS ABLE TO OPEN STUDY BEFORE COMPLETE STUDY IS TRANSMITTED TO VOXEL. THE DR. MARKS TREATMENT ISOCENTER ON THE INCOMPLETE STUDY AND SAVE DATA. LATER, AFTER TRANSMISSION IS COMPLETE, THE PATIENT STUDY IS OPENED AND THE MARKET ISOCENTER HAS SHIFTED BY THE DISTANCE OF ADDITIONAL AREA COVERED BY COMPLETE SCAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACQSIM | RADIATION THERAPY SIMULATION SYSTEM | KPQ | PHILIPS MEDICAL SYSTEMS | ACQSIM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |