FDA Adverse Event Malfunction Summary report: N

ACQSIM

MDR report key: 481461 · Received August 28, 2003

Report

Report Number
1525965-2003-00010
Event Type
Malfunction
Date Received
August 28, 2003
Date of Event
July 31, 2003
Report Date
July 31, 2003
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
KPQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ACQSIM-CT OPERATED WITH PREFERENCE OF AUTO SEND ON, TO SEND IMAGES TO VOXELQ. CUSTOMER IS ABLE TO OPEN STUDY BEFORE COMPLETE STUDY IS TRANSMITTED TO VOXEL. THE DR. MARKS TREATMENT ISOCENTER ON THE INCOMPLETE STUDY AND SAVE DATA. LATER, AFTER TRANSMISSION IS COMPLETE, THE PATIENT STUDY IS OPENED AND THE MARKET ISOCENTER HAS SHIFTED BY THE DISTANCE OF ADDITIONAL AREA COVERED BY COMPLETE SCAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACQSIM RADIATION THERAPY SIMULATION SYSTEM KPQ PHILIPS MEDICAL SYSTEMS ACQSIM NA

Patients

Seq Age Sex Outcome Treatment
1 *