1823260-2015-03576
Report
- Report Number
- 1823260-2015-03576
- Event Type
- Malfunction
- Date Received
- June 2, 2015
- Date of Event
- May 16, 2015
- Report Date
- June 2, 2015
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS EVENT OCCURRED IN (B)(6). (B)(4).
THE CUSTOMER RECEIVED QUESTIONABLE TROPONIN THS RESULTS FOR TWO SAMPLES FROM ONE PATIENT AND ONE SAMPLE FROM ANOTHER PATIENT. FOR PATIENT A: A TOTAL OF FOUR SAMPLES WERE DRAWN FROM THE PATIENT. THE DOCTOR QUESTIONED THE RESULT OF 36.64 (37) NG/L FOR SAMPLE 3 AND THE RESULT OF 3.90 (<5) NG/L FOR SAMPLE 4. REPEAT TESTING WAS THEN PERFORMED ON THESE TWO SAMPLES. FOR PATIENT B: A TOTAL OF FOUR SAMPLES WERE DRAWN FROM THE PATIENT. THE LABORATORY QUESTIONED THE RESULT OF 53.75 (54) NG/L FOR SAMPLE 2 AND REPEATED TESTING. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE PATIENTS WERE NOT ADVERSELY AFFECTED. THE REAGENT LOT NUMBER WAS 182603 WITH AN EXPIRATION DATE OF 04/29/2016. DURING INVESTIGATION AT THE SITE, PRECISION TESTING WAS ACCEPTABLE. NO RELEVANT HARDWARE ALARMS WERE FOUND EXCEPT ABNORMAL SAMPLE ASPIRATION ALARMS ON THE INITIAL TESTING OF SAMPLE 3 FOR PATIENT A. A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. ADDITIONAL INFORMATION FOR FURTHER INVESTIGATION WAS REQUESTED BUT WAS NOT PROVIDED. BASED ON THE PROVIDED INFORMATION, AN ISSUE WITH THE INSTRUMENT OR WITH THE CALIBRATION/REAGENT WAS UNLIKELY. A PRE-ANALYTICAL ISSUE WITH THE SAMPLES WAS MOST LIKELY THE CAUSE BASED ON THE ANALYZER ALARMS.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |